2. PANEL DESCRIPTION (Fig. A)
1. "-" Negative pressure hose plug.
2. HOLD Freezes pressure reading.
3. "-" Minus pressure display.
4. REC starts/displays max/min pressure.
5. MAX MIN pressure recorded.
6. BAT Low Battery Indicator
7. On/Off button
8. Hold button freezes pressure reading.
9. Unit button cycles through pressure readings.
10. "+" Positive pressure hose plug.
11. DIF Differential pressure mode.
12. Primary Data Screen.
13. H/M/S 88:88:88 Hr/Min/Sec.
14. Pressure unit indicator.
15. USB output port.
16. DC power input for AC adapter.
17. Rec button starts the relative clock.
18. Dif button displays relative zero.
19. Backlight button lights LCD for apx. 30 seconds.
20. Battery Compartment.
NOTE: Some functions shown on the display are not available.
3. INSTRUCTIONS
3-A. General Instructions
The Manometer measures “Gauge Pressure.” This combines Differential
pressure with a PSI measurement as referenced to ambient pressure. When
the manometer is first powered on, the entire LCD display flashes for
approximately 3 seconds. The LCD is divided into two distinct sections: One
large (Primary) top screen and one smaller section at the bottom-right of the
LCD (Relative Clock). The pressure measurements are constantly updated in
the 2 display areas. Check that the tubing is not leaking or damaged before
use. The start-up mode is programmable. The display will default to the mode
last used and the meter defaults to the setting used during the last operation.
 Turn the unit on with the ON/OFF button.
 Press the REC button once, and the relative clock starts in the lower right
screen. "REC" is displayed in the middle-left of the LCD (Pg 6, Fig B). The
DIF button will not operate at this time.
 Press the REC button repeatedly to cycle through MAX (Pg 6, Fig C) and
MIN (Pg 6, Fig D) and back to the current pressure. To exit the record
mode, press and hold the REC button for 3 seconds.
1
2
3
4
5/6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
-8.8.8.8
HOLD DIF
REC H/M/S
MAX MIN BAT
88:88:88